How to fix air and powder supply issues in NSK Medical Equipment?
- VVictoria HarringtonSep 10, 2025
If air and powder are not being supplied, or the pressure is weak in your NSK Medical Equipment, try these steps: * Replace the Nozzle Tip, as it may be clogged. * Clean the Nozzle base with the small Cleaning wire and the Nozzle end with the cleaning file. Also, clean the inside of the Handpiece with the large Cleaning wire. If this doesn't work, try Ultrasonic Cleaning with warm water. * Remove powder from inside the Chamber. * Perform Auto-Cleaning to remove any solidified powder in the air passage. * If the powder container has been left open, dispose of the powder and use a fresh supply. * Drain the compressor by opening the filter in the air supply circuit and check for water accumulation inside the air filter and compressor drain.
 