EasyManua.ls Logo

OHAUS DV114C - 5 Technical Data; Drawings

OHAUS DV114C
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN-43Ohaus Discovery
5. TECHNICAL DATA
Ambient conditions
The technical data are valid under the following ambient conditions:
Ambient temperature: 10 °C to 40 °C
Relative humidity: Max. 80 % at 31 °C linearly decreasing to 50% at 40 °C, non-condensing
Height above sea level: Up to 4000 m
Warm-up time: At least 120 minutes after connecting balance to the power supply
Storage temperature range: -10 °C to +70 °C
Voltage fluctuations: -15% + 10%
Power
AC adapter - To national standard as per list in Section 4.5. Balance power input
12 VAC, 50/60Hz 1.0A
Materials
Base Housing: Painted Aluminum
Draft Shield: Painted Aluminum
Rear Cover: Plastic (Polyurethane)
Terminal: ABS / PC Blend Plastic
Weighing Pan: Stainless Steel
In-Use Display Cover: PET Plastic
Draftshield Doors: Glass
Protection
Protected against dust and water
Pollution degree: 2
Installation category: Class II
EMC: see declaration of conformity
5.1 Drawings