SJX SERIES BALANCES EN-11
Span [SpaN]
Initiates a span calibration procedure (zero and span)
Lin [LIN]
Initiates a linearity calibration procedure (zero, mid-point and span).
End Cal [End]
Advance to the next menu or return to the top of the current menu.
4.3 Setup Menu
Enter this menu to set balance parameters.
Note: Bold always represents factory default value.
Reset [reset]
Reset the Setup menu to factory defaults.
NO = not reset
YES = reset
Filter [FiltEr]
Set the amount of signal filtering.
LOW = less stability, faster stabilization time
MED = normal stability, stabilization time
HI = greater stability, slower stabilization time
AZT [AZt]
Set the automatic zero tracking functionality.
OFF = disabled
0.5d = the display will maintain zero until a change of 0.5 divisions per second has
been exceeded.
1d = the display will maintain zero until a change of 1 divisions per second has been
exceeded.
3d = the display will maintain zero until a change of 3 divisions per second has been
exceeded.
Stable Range [Stable]
Set the amount the reading can vary while the stability symbol remains on.
0.5d = 0.5 balance division
1d = 1 balance division
2d = 2 balance division
5d = 5 balance division
Reset: no, yes
Filter: Low, Med, High
Auto Zero Tracking: off, 0.5d, 1d, 3d
Stable: 0.5d, 1d, 2d, 5d
Backlight: off, on, auto
Auto Tare: off, on, on-acc
Auto Off: off, 1, 5, 10
End Setup: Exit menu