EasyManua.ls Logo

OHAUS Valor 2000 - Troubleshooting Symptom Guide

OHAUS Valor 2000
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 2 DIAGNOSTIC GUIDE
Ohaus Corporation www.ohaus.com 2-6 Valor 2000 Series Service Manual
2.2.4 Diagnostic Guide
TABLE 2-4. DIAGNOSTIC GUIDE
Symptom
Possible Cause
Remedy
Unstable weight
Air flow may be blocked to
scale because rear feet are
tightened completely.
Adjust rear feet at least slightly to
allow proper air flow. (They should
be tightened when scale is washed.)
-----
Busy message
Displayed while zero or tare function
is in process
--C--
Busy message
Displayed while calibration data is
being collected.
CLEAr PAN
Clear pan message
Weight reading exceeds zero limit.
Remove load from scale.
dONE
Done message
Calibration process is completed.
-OFF-
Off message
Scale is turning off.
-NO-
No message
Requested action is not allowed.
loBat
Low Battery
Battery charge is too low.
Recharge battery.
CAL E
Calibration Error
Calibration point value is outside of
acceptable range. (See Appendix A
or B.)
Err 1
Error 1: Overload limit
exceeded
Weight reading exceeds capacity
plus 9d. Remove load from scale.
Recalibrate scale.
Err 2
Error 2: Underload limit
exceeded
Weight reading is below calibration
zero value by 30% of scale capacity.
Add load to scale. Recalibrate scale.
Err 9
Error 9: Menu data checksum
error
Menu data in EEPROM is corrupted.
Cycle power on and off. If error
continues, check menu settings.
If problem continues, replace PCB.
Err14
Error 14: Initial Zero limit error
Weight reading exceeds 10%
Power On Zero limit when LFT is set
to ON.
Place platform on the balance. If the
problem continues, add a small
amount of weight at power on. If the
problem persists, perform span and
linearity calibration from Service
menu. (See Appendix B.)

Table of Contents

Other manuals for OHAUS Valor 2000

Related product manuals