1.4 SPECIFICATIONS
Complete specifications for the Ohaus Valor 2000 Scales are listed in Table 1-2. When a scale
has been serviced, it must meet the specifications listed in the table. Before servicing the scale,
determine what specifications are not met.
TABLE 1-1. SPECIFICATIONS
Model
V21PW3
V21PW6
V21PW15
Capacity x Readability
Max x d
(non-approved)
3.0000 kg x .0005 kg
3000 g x 0.5 g
6 lb x 0.001 lb
96.00 oz x .02 oz
6.000 kg x .001 kg
6000 g x 1 g
15 lb x 0.002 lb
240.00 oz x .05 oz
15.000 kg x .002 kg
15000 g x 2 g
30 lb x 0.005 lb
480.0 oz x .1 oz
Maximum Displayed
Resolution
1:6000
1:7500
1:7500
Capacity x Readability
Max x d (approved)
3 kg x 0.001 kg
3000 g x 1 g
6 lb x 0.002 lb
96 oz x 0.05 oz
6 kg x 0.002 kg
6000 g x 2 g
15 lb x 0.005 lb
240 oz x 0.1 oz
15 kg x 0.005 kg
15000 g x 5 g
30 lb x 0.01 lb
480 oz x 0.2 oz
Lo-res Displayed
Resolution
1:3000
1:3000
1:3000
Linearity and Accuracy
±1d
Off Center Load error
±1d
Repeatability
±1d
Creep (5 sec 5 min)
1d
Hysteresis
1d
Zero drift vs. temp.
±1 d /
o
C
Span drift vs. temp.
± 1d / C
Zero Range
±2% of Full Scale Capacity
Tare Range
To Full Capacity by Subtraction beyond zero range
Stabilization Time
2 seconds
Overload Indication
Capacity + 9d
Warm-up Drift at Zero
(Over 5 min)
5d max
Operating Temp. range
to 40°C / 32° to 104°F
Storage Temp. range
-10° to 60°C / 14° to 140°F
Humidity Range
Maximum relative humidity 80 % for temperatures up to
31 °C / 88 °F, decreasing linearly to 50 % relative humidity
at 40 °C / 104 °F, non condensing.
Safe Overload Capacity
120% of capacity
AC Power
9VDC (700mA)