APPENDIX B SERVICE CALIBRATION
Valor 2000 Series Service Manual B-3 Ohaus Corporation www.ohaus.com
B.5 Service Span Calibration
When S.CAL appears, press YES to accept the Service Calibration menu,
or NO to advance to the End sub-menu selection.
S.CAL
When S.SPAN appears, press YES to accept the Service Span Calibration
menu or No to advance to the Service Linearity Calibration menu.
S.SPAN
0 flashes. With no weight on the pan, press Yes to establish the zero
point. (Outlined characters represent a flashing display.)
- - C - - flashes while the zero point is established.
--C--
Service Span Calibration point flashes.
(The example shows the value for a Scale Capacity of 3 kg.)
Place the specified calibration weight on the pan and press Yes.
- - C - - flashes while the span point is established.
--C--
If span calibration was successful, the actual weight reading appears for
three seconds, followed by the Service Linearity Calibration menu item.
(The example shows a 3 kg span weight on the pan.)
3.000