EasyManua.ls Logo

OHD Quantifit2 - Page 48

OHD Quantifit2
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48
Chapter 17
| Tests |
17
Tests
The testing sequence
o When initiating a test from Logic software:
·
The software will display the inhalation
valve message reminder.
·
Then the software will display the tube
connection instructions.
·
The previous two screens can be
turned off by checking the box
next to "Hide these messages next
time."
·
The test subject will view a
message that says, "Device is in
remote mode. Please wait until the
Operator sets up the test."
·
If the daily verification check has
not been completed for the day,
the QuantiFit2 will prompt the user
to perform this check.
·
When the test is initiated, the test subject will see "Please review this information. If
correct, press the blue arrow to continue." (Name, size respirator, parameter, surrogate)
·
Test subject will then follow on-screen instructions to complete the test.
o If a test step should fail for any reason
·
The operator will be notified on the software test screen, and be given an option to
retry the step or save it (if a fit factor was calculated).
·
There are some troubleshooting suggestions that can be seen by clicking on the
question mark in the upper left of the test screen next to the Delete button.
·
See the 20 Troubleshooting section for more information.
54

Table of Contents