Operation of CASY TT
MAIN MENU – Select Capillary
. MAIN MENU – Select Capillary
The Select Capillary option opens the CAP SELECTION screen that enables you to switch the
system to an alternative capillary size. Depending on how your system is congured, one or three
entries are available for dierent measuring capillaries.
CAP. SELECTION TT-2AB-9999 15.04.2009 10:15
Capillary: 150 µm 60 µm 45 µm
*
USE NEXT TO SELECT CAPILLARY AND PRESS ENTER TO LOAD.
USE CANCEL TO LEAVE SCREEN.
Pressing NEXT enables you to switch between the measuring capillary sizes available on your sys-
tem. The selected measuring capillaries are tagged with *. Press ENTER to load the calibration
for the selected measuring capillaries. The setup previously used with this measuring capillary is
activated. Please note that previously displayed measuring data are lost when a capillary setting is
changed. Pressing CANCEL enables you to exit the screen without changing the settings.
. MAIN MENU – Set Date and Time
Your CASY TT has an integral Real-Time-Clock. The Set Date and Time function opens the DATE
AND TIME screen, in which you can set date and time. You have the option of changing the date
display from EU format “DD.MM.YYYY” to US format “MM/DD/YYYY”.
Use the MENU key to move between settings and press NEXT to skip to the individual gures used
in the date and time. Use the
and to change the settings. Press ENTER to conrm your settings
or CANCEL to exit the screen without accepting any changes.
DATE AND TIME TT-2AB-9999 15.04.2009 10:15
Date Format: dd.mm.yyyy
Date: 15.04.2002
Time: 10:15:00
USE MENU AND NEXT TO STEP. USE +/- TO EDIT.
USE CANCEL TO LEAVE SCREEN OR ENTER TO CONFIRM CHANGES.