40
CASY TT Operator‘s Guide
Numerical Measuring Results
. Numerical Measuring Results
Depending on the Graph Parameters set on your system (see Chapter 5.7) you will nd the most
important measuring results at the top of the size distribution. You will also see, in the top right-
hand corner, the current setting of the evaluation cursors CL and CR and the normalization cursors
NL and NR, which are used as a basis to calculate the measuring results.
Pressing NEXT enables you to display the entire numerical evaluation of the measurement data.
Pressing NEXT again will return you to the size distribution.
SET : TUMOR-03¦DIL : 5.000E+01 ¦CAP : 150 µm
DATE: 15.04.2009¦SVOL: 3 x 400 µl¦SCAL: 40 µm
TIME: 10:15¦AGG : MANUAL ¦NORM: 7.60-40.00 µm
SN : TT-2XX-9999¦PVOL: 4.249E+03  ¦EVAL: 12.80-40.00 µm
--------------------------------------------
Counts : 2911 ¦ConcRng: OK
>###µm/ml: 1.500e+02 ¦Aggreg : 1.157
Viable/ml: 1.684e+05 ¦%Viable: 83.04 %
Total/ml : 2.028e+05 ¦
Debris/ml: 7.483e+04 ¦
Volume/ml: 1.431e+07 ¦
MeanVol : 4.917e+03 ¦MeanDia: 21.10 µm
PeakVol : 4.250e+03 ¦PeakDia: 20.10 µm
The numerical results display is divided into two distinct areas. The header, which is separated by
a horizontal line, contains a complete documentation on all measuring parameters used for the
measurement. The main section contains all the measuring results for the set cursor ranges.
This example shows the format of the numerical display for settings %Calculation = %Via and
Aggr.Correct = Manual. Details on possible settings can be found in Chapter 5.4 Depending
on your settings, some of the parameters visible in this example may not appear or alternative
parameters are shown.