48
CASY TT Operator‘s Guide
.
Manual Cursor Control using Control Panel Keys
You can use the manual cursor control keys located on the left-hand side of the control panel to
change the position of the cursor while visually monitoring the size distribution. If you are in the
numerical results display or menu control, you must rst move to the display.
Using the L-CURSOR or R-CURSOR key activates the mode to manually set the cursor. This mode
is identied by the presence of an asterisk (*) in the top right-hand corner of the screen, just before
the cursor position to be adjusted using the arrow keys. The evaluation cursor setting is always ac-
tive after starting manual cursor control.
It is possible to alternate between L-Cursor and R-Cursor to activate the setting for the left and/or
right-hand evaluation cursor. You can switch to the normalization cursor setting by pressing NEXT.
Now you can use the L-Cursor and R-Cursor keys to activate the adjustment for the left and right-
hand normalization cursor respectively. Pressing NEXT again returns you to the evaluation cursor
setting.
The
and keys move the active cursor right and left respectively. The selected cursor
moves across the size distribution accordingly. At the same time, the numerical display of the cursor
position is constantly updated in the top right-hand corner of the size distribution. Pressing L-Cur-
sor, R-Cursor and NEXT enables you to select a dierent cursor to move at any time.
The manual cursor setting mode can be terminated by pressing either ENTER or CANCEL. If you
press ENTER, the new cursor positions are accepted as active settings and the calculation of the
displayed measured values is updated. If you press CANCEL, the previous cursor position settings
are restored. All new changes are rejected.
While in the manual cursor control mode, only the L-CURSOR, R-CURSOR, ,
, CANCEL and ENTER keys are active. You must leave the manual cursor control mode
before using the other control panel functions.
. Outputting the Results
The data can be output using either a parallel or a serial interface. The parallel interface is provided
to connect a PCL-capable printer (PCL = Printer Control Language). The data can be transferred to
a computer for further editing via the serial interface (see Appendix 8.1, Data Output via the Serial
Interface).
To export size distributions and measuring results into MS-Excel (PC) through a fully automated
process, you can use CASY excell which is delivered with each CASY TT System.
If you have set the Print Mode parameter to Auto, the data are automatically outputted after each
measurement via interfaces specied in the current parameters (see Chapter 5.4). If Print Mode is
set to Manual, you must start to output the measuring results by pressing PRINT.
Manual Cursor Control using Control Panel Keys