Phenomenon
Cause
Remedy
1
el
Resolution problems:
LB
objectives
other
than
Use
Olympus
LB
series objec-
series are used.
tives.
I
mage
is
not
sharp.
objectiv8
not
correctly
Click
nosepiece
into
place.
I
IS
Insufficient
contrast.
positioned in
the
light path.
Image details lack defi-
nition.
Objective correction collar
is
Rotate
correction
collar, keep-
not
adjusted
correctly.
ing
specimen in fine
focus
until
optimum
resolution
is
obtained.
Dust on objective
front
lens
Clean
front
lens.
Immersion objective
is
not
Use
immersion
oil.
used
with
immersion
oil.
Bubbles in immersion oil
Remove bubbles
(and
reap-
ply
oil).
Immersion
oil designated
by
Use
Olympus
immersion oil.
Olympus
is
not
w~ed.
Dirty
specimens
Clean.
Dust on condenser
lens
fl
Field
of
view
is
partially
Objective
is
not
correctly
Slightly
rotate nosepiece
until
out
of
focus,
or
image
is
positioned in the
light
path.
it
clicks in place.
partly
out
of
focus.
Specimen
is
not
correctly
Place specimen slide
correctly
positioned on stage.
on stage, and place
stage
clips
open
it.
gl
Specimen image
is
partial-
Objective
is
not
correctly
Slightly
rotate nosepiece
until
Iy
out
of
focus.
positioned in the
light
path.
it
clicks
into
place.
Condenser
is
not
centered.
Center condenser.
h)
Field
of
view becomes
Condenser
is
not
correctly
Center condenser.
only
slightly
brighter
by
centered.
increasing voltage.
Condenser
is
lowered
too
Adjust
condenser height.
much.
2.
Electric System
al
Illuminator
is
too
bright
Line
voltage
selector switch
Match
selector switch
to
(or
too
dark)
even
when
is
not
matched
with
local
mains voltage.
adjusting
control
lever.
mains voltage.
b)
Voltage
for
illuminator
cannot
be
raised.
cl
Lamp
goes
off
and on.
Bulb
filament
is
likely
to
Replace bulb.
burn
out.
Loose electric connections.
Check all connections.
d)
Bulb burns
out
frequently.
Line
voltage selector switch
is
Match
selector switch
to
not
matched
with
local mains mains voltage.
voltage.
Bulb
is
not
standard one.
Use
standard bu lb.