CX21 MAINTENANCE MANUAL INSPECTION PROCEDURE
- 8 -
CHAPTER 2
INSPECTION PROCEDURE
1. Checking Performance of Microscope
Using the CX21 inspection sheet (P.7), check the electrical unit, mechanical and
optical
performance.
2. Checking Dirty Portion
2-1 Image influence caused by dirt on each component
The following figure shows the influence of image on each optical component if stains
or dust is adhered to that portion.
In general, the microscope image is largely affected by dirt adhered on the nearer
portion to a specimen and image surfaces.
Therefore, the optical components should be kept clean and dust-free.
A: Dirt is clearly seen.
B: Blurred contours of dirt is seen.
C: Dirt is seen when the aperture iris diaphragm is stopped down.
D: Dirt is not directly seen, but contrast of image deteriorates.
D Objective
C Prism
A Specimen
B Condenser lens
C Frosted glass
B or A Glass
(Frosted : underside)
B Eyepiece
C Collector lens

Other manuals for Olympus CX21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus CX21 and is the answer not in the manual?

Olympus CX21 Specifications

General IconGeneral
TypeBiological Microscope
Power SupplyAC 100-240V, 50/60Hz
EyepieceWidefield 10x
Objective Lenses4x, 10x, 40x, 100x
NosepieceQuadruple
StageMechanical stage
FocusCoaxial coarse and fine focusing
IlluminationLED
CondenserAbbe condenser NA 1.25
Optical SystemUIS2 optical system
Eyepiece TubeBinocular

Related product manuals