CX21 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
- 2 -
2. Guide to Maintenance
2-1 Overview of maintenance
1)
OLYMPUS
CX21
100
110
120
130
140
150
17
160
0
10
20
30
Set your correct interpupillary distance.
Note any areas suggesting a need for
mechanical and/or optical maintenance
by operating it or observing a specimen.
2)
Sweep off dust on the outer surfaces
with the soft brush. If there are stains on
the microscope frame, clean them with
neutral detergent.
3)
Remove the eyepieces and objectives,
then clean their lenses with cleaning
solution. (When removing the objective,
use the rubber band.)
4)
Remove the condenser downward.
Wipe off any dirt spots on the surface of
top lens, filter. The plastic parts should
be cleaned with netural detergent.
5)
Take off the mechanical stage from the
plane stage by removing two screws.
Clean the surface of stage and filter on
the base cover.
6)
Mount the removed components and
mechanical stage to the frame.
7)
Polish all plastic components and the
frame with silicon cloth.
8)
Return the interpupillary distance to
original condition and prepare for the
inspection. (see page 6)
9)
Do a final check following the inspection
sheet in this manual. (See page 7)
Tool required:
Lens tissue Neutral detergent
(For plastic part or
frame)
Cleaning solution
(For lens or filter
etc.)
Cotton swab
or tweezers
etc.
Blower Silicon cloth
(For finishing)
OLYMPUS
CX21
100
110
120
130
140
150
17
160
0
10
20
30
OLYMPUS
CX21
100
110
120
130
140
150
17
160
0
10
20
30
OLYMPUS
CX21
100
110
120
130
140
150
17
160
0
10
20
30
Top lens
(glass)
Filter
( plastic )
OLYMPUS
CX21
100
110
120
130
140
150
17
160
0
10
20
30
OLYMPUS
CX21
100
110
120
130
140
150
17
160
0
10
20
30
75
70
60
50
OLYMPUS
CX21
75
70
60
50
100
110
120
130
140
150
17
160
0
10
20
30
OLYMPUS
CX21
75
70
60
50
100
110
120
130
140
150
17
160
0
10
20
30
Remove it from groove.
(A)
Turn knob (A).
Plastic
Plastic
Loosen the screw securing eyepiece.
glass