Michael ValdezSep 9, 2025
Why does the stage drift down on my Olympus CX41?
- FFrank SmithSep 9, 2025
If the stage drifts down by itself or focus is lost during observation with your Olympus Microscope, tighten the tension adjustment lever.

Why does the stage drift down on my Olympus CX41?
If the stage drifts down by itself or focus is lost during observation with your Olympus Microscope, tighten the tension adjustment lever.
What to do if coarse adjustment will not go all the way up on Olympus CX41 Microscope?
If the coarse adjustment will not go all the way up on your Olympus Microscope, unlock the pre-focusing lever.
What to do if coarse adjustment will not go all the way down on Olympus CX41?
If the coarse adjustment will not go all the way down on your Olympus Microscope, raise the condenser holder.
Why does the objective hit the specimen before focusing on Olympus CX41 Microscope?
If the objective makes contact with the specimen before focus is obtained on your Olympus Microscope, mount the specimen correctly.
| Type | Biological Microscope |
|---|---|
| Optical System | UIS2 Optical System |
| Objectives | Plan Achromat objectives (4x, 10x, 40x, 100x Oil) |
| Nosepiece | Revolving nosepiece for 4 objectives |
| Power Supply | AC 100-240V, 50/60Hz |
| Illumination | 6V/30W halogen bulb |
| Focusing | Coaxial coarse and fine focusing knobs |
| Stage | Fixed stage with specimen holder |
| Eyepiece Tube | Binocular or Trinocular |
| Condenser | Abbe condenser, NA 1.25, with iris diaphragm |
| Eyepiece | 10x/20mm high eyepoint, widefield (depending on model) |