Timer Setting:
(Found on the back of the CO² regulation unit)
The endoscopic CO² regulation unit has a built in timer, and will automatically stop feeding
the gas as the set time elapses.
See below the available timer settings.
Select OFF to negate the automatic gas function
____________________________________________________
Timer setting Time before gas stop
_______________________________________________________________
Long About 30 minutes
_______________________________________________________________
Short About 15 minutes
_______________________________________________________________
OFF The Automatic stop function is turned off
_______________________________________________________________
Slide the timer switch to select the timer setting