9. Lamp will ignite automatically or to ignite light lamp press and hold lamp “Standby
switch” until it beeps”. Lamp turns on and green light next to “ON” illuminates”.
10. “White Balance Incomplete” may appear on the screen. To white balance endoscope,
place distal tip of scope into white balance barrel (cap). Press and hold down the “White
Balance” button until you hear a beep and the words “White Balance complete” appears
on the screen.