• The problem may be caused by foreign particles attached to the electrical contacts. Take the following steps and then
connect the connector to the light source.
Turn off the processor.
Moisten lint-free gauze with 70% ethanol. Wipe electrical contacts and leave them to
dry naturally.
Turn off the light source. Detach the scope connector.
1 2
4 5
3
What to do if abnormal images appear after turning on the power
• The foreign particles might have accumulated on the light source output socket or other problems may cause abnormal
images.
• Please contact Olympus for further troubleshooting assistance.
Normal condition inside the output
socket.
Accumulated lint inside the output
socket after connectors with lint were
repeatedly connected.
What to do if abnormal images still appear even after taking the above steps
QR 07.281 V1.0 November 2019
For complete product details see Instructions for Use.