Chapter 6 Measurement procedures
Scaler operating procedure
1 Startup
Display the image to be measured and press the [MEAS/ENTER] joystick.
2 Reference point specification
Place the cursor on each extremity of an object with a known length in the
observed image and press the [MEAS/ENTER] joystick.
3 Reference length specification
When the reference length input message is displayed, input the known length
by tilting the [MEAS/ENTER] joystick up/down and left/right, and press the
[MEAS/ENTER] joystick to enter it.
4 Measurement point specification
Place the cursor on each extremity of the measurement object and press the
[MEAS/ENTER] joystick to measure its length.
5 Exit
Press the [LIVE/GAIN] button to display the exit measurement screen and then
select “OK” to exit measurement. Select “CANCEL” to return to the stereo
measurement screen.
Select “EXIT” in the scalar measurement menu to stop measuring and return to
the screen before measuring was activated.
Specifying the
measurement point