4.5 Adjusting the image
Still image (freeze)
CAUTION
Do not insert and withdraw the scope when the image is frozen.
1 Press the [FRZ/REC] button on the side of the control unit to freeze
the observed image. The LCD monitor displays the freeze indicator
( ) on the top right of the screen.
NOTE
If the [FRZ/REC] button is held depressed for more than 2 seconds, image recording will be
initiated.
2 When the observation image is frozen, press the [FRZ/REC] button
again or the [LIVE/GAIN] button to return to a live image.
3 When an image containing quick action is frozen, the frozen image
may be blurred. Further, when the “NOISE REDUCTION” setting
(page 94) is “HIGH”, ghosting is more likely to appear. Change the
setting to “STANDARD” to reduce ghosting.