INSTRUCTIONS
IV9000N
IV9435N
IV9450N
IV9635N
IV9635X1N
IV9650N
IV9675N
INDUSTRIAL ENDOSCOPE
IPLEX NX
Industrial Endoscope
MM2671 08 Issued in February, 2020

Table of Contents

Questions and Answers

  • S
    Sharon WestSep 23, 2025
    Why doesn't the stereo measurement function start on my Olympus Industrial Equipment?
    • P
      powellvanessaSep 23, 2025
      If the stereo measurement function does not start, verify that the combination of the stereo optical adapter and the stereo scope unit is correct by checking their serial numbers.
  • R
    Rachel HernandezAug 25, 2025
    What should I do if my Olympus IPLEX NX IV9435N displays
    • J
      Joel SheaAug 25, 2025
      If your device shows this message, it means the SDHC card or USB flash drive is full. You should delete any unnecessary data to create more space, or replace it with a new SDHC card or USB flash drive.
  • C
    cassandranicholsAug 25, 2025
    Why is my Olympus IPLEX NX IV9435N image not sharp?
    • A
      Andrew NortonAug 26, 2025
      The lack of sharpness may be due to a dirty objective lens on the distal end of the insertion tube or the optical adapter. Try wiping it with a clean piece of gauze or a cotton swab. Also, ensure the optical adapter is correctly attached and that both the image sharpness and external monitor are properly adjusted.
  • K
    Kristy MatthewsAug 26, 2025
    What to do if Olympus IPLEX NX IV9435N displays
    • C
      Charles BarnesAug 27, 2025
      This message indicates that the battery voltage is low. Stop all operations like recording, copying, deleting, or formatting the USB flash drive. Replace the battery immediately or connect the AC adapter.
  • S
    Sylvia GriffinSep 23, 2025
    Why is the measurement result of the scaler measurement not displayed on my Olympus IPLEX NX IV9435N?
    • B
      Beth MartinezSep 23, 2025
      If the measurement result of the scaler measurement is not displayed, it may be due to using an image acquired by an unsupported software version (1.20 or before). Use an image acquired by software version 1.20A or later.
  • J
    Jasmine McintyreSep 13, 2025
    What to do if Olympus IPLEX NX IV9435N displays
    • A
      Amanda Mccullough MDSep 13, 2025
      Only images recorded with this instrument can be displayed.
  • T
    thoodSep 12, 2025
    Why does my Olympus IPLEX NX IV9435N Industrial Equipment optical adapter not attach to the insertion tube?
    • C
      Corey NelsonSep 12, 2025
      The inability to attach the optical adapter could be due to a foreign object stuck to the screws. Try wiping with a clean piece of gauze or a cotton swab. It could also be that you are not using the designated optical adapter for this system, so ensure you are using the correct adapter. Finally, the installation procedure may be incorrect.
  • M
    Marcus HarrisSep 12, 2025
    Why won't my Olympus IPLEX NX IV9435N Industrial Equipment turn off?
    • D
      Diane BoyleSep 12, 2025
      If the system cannot be turned off, the main unit might be damaged. Remove the AC adapter or the battery from the main unit and turn OFF the power.
  • D
    Donna CarrSep 4, 2025
    What to do if Olympus IPLEX NX IV9435N displays
    • M
      Maria CampbellSep 4, 2025
      This message indicates that the self-check function was activated due to an overcurrent in the light source. Terminate the inspection, allow the instrument to cool down, and then turn the power back on again.
  • F
    floresemilySep 2, 2025
    What to do if Olympus Industrial Equipment displays
    • A
      Alejandro ChristensenSep 2, 2025
      This message indicates that the distal end has become too hot, activating the self-check function. Immediately pull the insertion tube out of the observation object.

Summary

Safety Precautions

General Safety Guidelines

Essential precautions for safe handling and use of the instrument.

Hook Assembly Safety

Specific safety measures for handling and attaching the hook assembly.

Battery Handling Safety

Precautions for safe use, charging, and disposal of the instrument's battery.

Nomenclature

Pre-Operation Preparation and Inspection

Power Supply Preparation

Instructions for loading the battery and connecting the AC adapter.

Pre;Post-Operation Inspections

Checklists for ensuring instrument integrity before and after use.

Basic Operation

Powering On and Illumination Control

Steps to turn the instrument on and adjust illumination levels.

Instrument Operation and Observation

Guidance on using controls, observing objects, and withdrawing components.

Image Recording and Playback

Procedures for capturing, saving, and reviewing images and movies.

Menu Operations and Functions

Stereo Measurement Function

Checking Measurement Values

Verifying measurement accuracy before and after the procedure.

Stereo Measurement Process

Detailed guide on performing measurements and recording results.

Scaler Measurement Function

Scaler Measurement Procedure

Detailed guide on setting reference points, length, and recording results.

Troubleshooting

Troubleshooting Guide and Error Messages

Guidance on identifying and resolving instrument issues and error indicators.

Common Problems and Solutions

Addressing frequently encountered issues and their recommended fixes.

Storage and Maintenance

Olympus IPLEX NX IV9435N Specifications

General IconGeneral
ModelIPLEX NX IV9435N
CategoryIndustrial Videoscope
Resolution640 x 480 pixels
Light SourceLED
Working Length3.5 m
Field of View90°
Weight2.9 kg
Storage Temperature-20 °C to 60 °C
IP RatingIP55
Image SensorCMOS
Insertion Tube Diameter4.0 mm
Direction of ViewForward