8
CARE AND STORAGE
8-1 Care After Use
Disconnect the A-cord, S-P cord and footswitch from the PSD-10.
Lightly wipe all surfaces using a soft cloth or gauze sponge. If dirt remains,
moisten the gauze with disinfectant ethanol and wipe again.
Wipe cords clean.
Wipe panel clean.
8-2 Storage
Band
For disinfection, wipe with a gauze sponge moistened with disinfectant ethanol
or 2% glutaraldehyde solution. If a glutaraldehyde solution is used, wipe
again with disinfectant ethanol to remove all residue.
Avoid touching the electrical contacts and connectors: poor electrical contact
will result.
To prevent scratches, do not use harsh or abrasive wiping material.
Surfaces must be thoroughly dry before use.
Wind the power cord around the hanger on the rear of the PSD-10.
Place the PSD-10 on a stable surface in the horizontal position.
Put on the vinyl dust cover.
Storage area must be maintained at normal temperature and humidity,
and away from water splashes.
Do not subject to heavy blows or impacts.
Put on the vinyl cover.