80
Appendix
ELECTROSURGICAL UNIT PSD-30
Item Specification
High
frequency
output
Output control Control by foot switch
Output time 10 seconds ON and 30 seconds OFF output (to
continue the output, keep the output time below
10 seconds)
Output power
vs setting
Rated load resistance 300
25 50
SETTING (W)
0
25
50
PURE
25 50
SETTING (W)
0
25
50
BLEND1
25 50
SETTING (W)
0
25
50
BLEND2