8-12
STRAIN
0
8
Scaling to
Display
Engineering
Units
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted)
INP.3=0 Unipolar input
(S2B jumper omitted)
INP.4=0 Std, for current inputs
INP.5= Disregard not used
INP.6=0 Disables IN.SC.OF
INP.7=1 Ratiometric input
IN.SC.OF Skip past
DEC PT FFFFF.F By shifting the decimal
point, it makes the meter
10 times more sensitive
(desired decimal point
will be entered later)
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
2. Now you are in the run mode and the load cell is connected to
the meter. Simulate two loads on the transducer and record the
meters display as “INPUT1” and “INPUT2” and record what
each of these readings (“READ 1” and “READ 2”) should display
in the spaces provided below. These numbers will be the INPUT
values entered in the “IN.SC.OF” scaling. Normally the two
points that are calibrated are the no load condition and 3/4 full
scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load the display indicates “–55.6”,
and when you simulate 4.0000 pounds the display indicates
“16136.6”.