Section 6: Battery InformationSection 6: Battery Information
Battery Replacement
This product does not have any user-serviceable parts except the battery which should be replaced periodically. The bat-
tery life is affected by battery type, ambient temperature, sample rate, sensor selection, off-loads and LCD usage. The
device has a battery status indicator on the LCD. If the battery indication is low, or if the device seems to be inoperable, it
is recommended that the battery be replaced.
To replace the battery, locate the battery compartment cover on the back of the unit. Remove the two (2) 3/32 hex
screws and expose the battery compartment. Use the pull tab to remove battery from compartment. Remove the old 9V
battery from the battery clips and replace with a new 9V battery. We recommend using 9V lithium battery. An alkaline
battery is acceptable, but will yield a shorter battery life.
BATTERY WARNING
This data logger contains a lithium battery. Do not cut the battery open, incinerate, or recharge. Do not heat
lithium batteries above the specifi ed operating temperature.* Dispose of the battery in accordance with local
regulations.
*See the individual specifi cation sheets at www.omega.com.