Data Logger Manual
Page 5
2.3 Specifi cations
Internal Temperature Channels
Range: -20 to +60°C
Resolution: 0.1°C
Accuracy: ±0.5°C (0 to +50°C)
Thermocouple Channels
Types: J, K, T, E, R, S, B, and N
Connection: Female sub-minature (SMP)
Cold Junction Compensation: Automatic, based on internal channel
Maximum Thermocouple Resistance: 1000Ω < 100Ω recommended
Thermocouples:
Type Range Resolution Accuracy
J -200 to +760°C 0.1°C ± 0.5°C
K -200 to +1370°C 0.1°C ± 0.5°C
T -200 to +400°C 0.1°C ± 0.5°C
E -200 to +980°C 0.1°C ± 0.5°C
R -50 to +1760°C 0.1°C ± 2.0°C
S -50 to +1760°C 0.1°C ± 2.0°C
B +50 to +1820°C 0.1°C ± 2.0°C
N -200 to +1300°C 0.1°C ± 0.5°C
Dot-Matrix LCD
Dimensions: 2.5” x 1.375” (63mm x 35mm)
Text: Confi gurable channel text size
Indicators: Power, status, memory
Backlight: Confi gurable with auto shut-off and contrast adjustment
Start/ Stop Modes
Software programmable start time and date, up to six months in advance, or manual (push-button) start.
Programmable stop time.
Memory
500,000 readings per channel with all channels enabled; 4,000,000 total readings; software confi gurable memory
wrap
Reading Rate
4 readings per second (4Hz) up to 1 reading every 24 hours