button on the CPC 100 front panel, and keep it pressed as long as an output
signal is not absolutely necessary for the test.
When connecting cables to the front panel input/output sockets, use wires
with 4 mm safety "banana" connectors and plastic housing.
For the connection between the CPC 100 and CP SB1 only use the specially
manufactured cables supplied by OMICRON electronics (see 1.4 "CP SB1
Accessories" on page 17).
Note: One end of the high-voltage cable has a coaxial safety plug that is
certified for a voltage level of 2 kV AC. The other end is equipped with a
safety banana plug that is insulated with a shrink tube.
When the CPC 100 is activated, consider this part of the cable to be a hazard
of electric shock.
Do not stand right next to or directly underneath a connection point because
the clamps may fall off and touch you. This is a physical and an electrical
hazard.
The red warning light on the CPC 100 front panel indicates hazardous
voltage and/or current levels at the CPC 100 outputs (red light "I" on or
flashing). The green warning light indicates that the CPC 100 outputs are not
activated.
Note: If none or both warning lights are on, the unit is defective and must not
be used anymore.
Static Charges
Static charges on transformer windings may be induced by test potentials. While
the voltage may not be significant enough to cause any damage, it can be a
source for serious accidents due to falls caused by reflex action.
Connect the windings to ground as described in the CPC 100 Reference
Manual.
Orderly Measures
The CP SB1 User Manual or alternatively the e-book in PDF format has always
to be available on site where the CP SB1 is being used. It must be read and
observed by all users of the CP SB1.
The CP SB1 may be used only as described in 3 "Application" on page 21. Any
other use is not in accordance with the regulations. The manufacturer and/or
distributor is not liable for damage resulting from improper usage. The user
alone assumes all responsibility and risk.