5. Maintenance
5-1. Routine Maintenance
Regular maintenance will maximize the service life of the device. Routine maintenance procedures
and supplies are explained below.
Maintenance Procedures
Main Unit
To clean the surface of the main unit, use a cloth moistened with a neutral detergent or disinfectant
alcohol.
Cuffs
To clean parts that come in direct contact with the skin, use a cloth moistened with a diluted
disinfectant alcohol. For other parts, simply remove foreign objects and dust; do not wipe with
alcohol or water.
Warning
After cleaning the device, dry it completely before turning on the power again. Otherwise
electric shock or current leakage may result.
Use only the specified supplies for the cord, cuff, and other parts. Do not install other
than specified options. An accident may result.
Caution
Before cleaning or maintaining the device, disconnect the power plug. Otherwise
electric shock or injury may result.
The ECG clip electrode and PCG sensor pad are disposable supplies. Do not reuse
them once they are removed. In case that they are used on moist, injured or infected
skin, dispose them right away. Otherwise an infection may result.
Note:
Do not wipe the connectors.
Note:
Dispose dirty or old cuffs as medical waste (do not recycle).