12
1-4. Safety Information
Important:
Before use
Make sure that all cords and tubes are firmly connected.
Inspect dials and buttons and verify that the device operates correctly.
Check printer toner, paper, and other supplies.
Confirm the expiry date of the ECG clip electrode or the PCG sensor pad.
The expiry date can be calculated from the lot number as follows:
During use
Do not pressurize when the cuff is not wrapped.
Do not use a torn cuff.
The device may be used on the patients below. If used on patients other than those described
below, incorrect measurement may result or measurement may not be possible.
- Height: 120 to 210 cm
- Girth of arm: 20 to 32 cm (using the standard cuff) / 16 to 38 cm (using the optional cuff)
- Girth of ankle: 16 to 33 cm
Do not disconnect the USB or LAN cable while data is being transmitted. This may corrupt the
data.
Do not turn off the power during printing, data transmission, or writing.
Do not pull out the paper before printing is finished.
Make sure the date and time setting is correct. If the date and time setting is not correct, the date
and time of measurement will not be recorded correctly.
Do not connect an ECG clip to another conductive part including the ground.
The electrical energy from a defibrillator may damage the device. During defibrillation, remove the
sensors from the patient and do not touch the device.
When moving or using the device, do not drop or subject the device to shock. This may cause the
electric components and precision mechanisms to fail.
After use
Clean the device and accessories and arrange properly for storage.
Do not wash or moisten the cuff
Do not use solvents such as thinner, benzene, or concentrated alcohol to clean the device
Do not use an autoclave, ultraviolet radiation, or gas disinfection (EOG, formaldehyde gas,
concentrated ozone, etc.) to disinfect the device
If a fuse blows, there may be a problem in the device. Contact a dealer or an Omron Healthcare
technical support representative
Do not install the unit in the following locations
- A location with prolonged exposure to direct sunlight
- A location with dusty or salt air
- On an inclined surface or a location subject to vibration, shock, or noise
- A location where chemicals are stored or gas is emitted
If the patient has one of the following conditions, a correct measurement may not be obtained
- The patient has body movements due to convulsions caused by rheumatism or otherwise
- The patient has diabetic arteriosclerosis (blood pressure at leg joints tends to be high)
- The patient has false high blood pressure
- The patient has convulsions or tremors
When the unit is incapable of ECG measurement, the message "Electrode removed" is displayed.
Lot Number:
Production Date: 2006 July 10th
Expiry Date: 2008 July
6071011
(2 years after the production date)
(Example)