35
2-1. Preparing for Measurement
Setting the Measurement Sites
The cuffs are normally attached to both brachia and both ankles.
If a shunt is placed on an arm for hemodialysis, do not attach a cuff to or perform measurement on
that arm.
1. Press the [MEAS. SITE] button.
2. Set the measurement sites.
Specify settings for the right brachium and left
brachium.
Cuff attached: [ON]
Cuff not attached: [OFF]
Specify settings for the right ankle and left ankle.
Ankle cuff attached: [ANKLE]
Toe cuff attached: [TOE]
Cuff not used: [OFF]
[TOE] only appears when the TBI package is installed.
3. Press the [OK] button.
To cancel entry, press the [CANCEL] button.
Notes:
When a TBI package is installed, read the accompanying manual.
The right brachium and left brachium cannot both be set to "OFF".
Measurement cannot be performed with both an ankle cuff and a toe cuff attached.