9 Maintenance
9.1 Preventive maintenance
Note!
The used detector is not subjected to measurable aging when used properly.
Tab. 18 Preventive maintenance
Component Maintenance activity Maintenance interval Information
Wetted parts
inspection with regard
to leakages
as part of standard
installation
maintenance
Possible damaged sealing faces
of the sensor body can lead to
leakages (see instruction manual
of the armature).
Lamp
module
exchange 1 - 2 years
Operation of the lamp below its
nominal voltage (4.8 V DC
instead of 5.0 V DC) enhances
lamp life. Strong vibrations, high
temperatures or frequent on and
off switching of the system can
have negative effects on the
service life. Statistical service life
is 3 years.