EasyManua.ls Logo

Optomed Aurora - Page 31

Optomed Aurora
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Optomed Aurora 31
Important information |
Optics module not detected The camera is not able to detect the optics module.
Please detach the optics module and restart the camera.
Patient list upload failure This indicates that there is a problem with patient list
uploading. Please retry. If the problem persists, please
contact customer service.
Unable to create new study The patient already has five studies and it is not possible to
create a new study. Copy images to a safe location and select
‘Erase image memory’ from Camera settings.
Charging failure This indicates that Aurora Camera is not properly
connected to Charging Station. Please reconnect the
camera to Charging Station.
Optics communication failure This indicates that there is a problem with the optics module.
Please reattach the optics module.
Autofocus calibration needed Optomed Aurora needs to be calibrated. Please hold
the camera horizontally and press Optoroller to calibrate.
Memory erasing failed This indicates that there is a problem with memory erasing.
Please retry. If the problem persists, please contact customer
service.
System memory not available This indicates that there is a problem with the system
memory. Please restart the camera.
Image counter full Aurora Camera can store 9,999 images. Once the image
counter is full, copy the images to a safe location and select
‘Erase image memory’ from Camera settings.
If the device behaves abnormally, Aurora Camera can be forced to shut down by pressing the power and
back buttons simultaneously for seven seconds or longer.
The manufacturer of Optomed Aurora will provide soware updates as needed throughout the lifecycle of
Optomed Aurora to continue to guarantee its safety and eectiveness.
NOTE!
If a malfunction cannot be rectified, please contact your local distributor or Optomed customer
service.

Other manuals for Optomed Aurora