Optomed Aurora 33
6.4 Service and repair
The user of the device is allowed to replace the battery and clean the device, which are instructed in
this user manual. There are no other maintenance procedures that can be carried out by the user. All
servicing and repairs other than replacing the battery and cleaning the device must be carried out by
Optomed or Optomed-authorized service facilities and service personnel. Optomed will make available
work instructions to repair those parts of medical electrical equipment that Optomed has designated as
repairable by service personnel.
If your device requires a warranty, extended warranty or non-warranty repair service, contact Optomed
Customer Service (service@optomed.com) or your local distributor. Estimates for non-warranty repairs
are provided at the currently valid charge; however the device must be sent to Optomed for an estimate.
When you contact customer service, the representative will record all necessary information and will
provide a Return Authorization Number. Prior to returning any product for repair, a Return Authorization
Number must be obtained.
6.5 Battery
CAUTION!
Only use the battery provided by Optomed with this product. Do not use a damaged or leaking
battery. Do not disassemble, modify, crush or destroy the battery pack. Doing so can cause
battery fluid leakage, heat generation, burns, fire and/or explosion.
CAUTION!
Charge the battery with Aurora Charging Station only. Use of an unrecommended charger may
cause battery fluid leakage, overheating of the battery, or may cause the battery to explode.
Aurora Camera has a rechargeable Li-ion battery that is charged when the camera is placed on Aurora
Charging Station, which is connected to the mains. Aurora Charging Station can also be used as an external
battery charger for the spare battery included in sales case. The battery pack is specially designed and
manufactured for this device. Optomed Customer Service or your local distributor provide suitable battery
packs. The label on the battery includes the following information:
Li-ion Rechargeable Battery 3.63 V / 2600 mAh / 9.438 Wh
50000655, 1ICR19/65, ICR18650-26J, WWYY (WW is week and YY is year of manufacture)
Provided for Optomed Oy by Celltech Oy, Finland
Li-ion Rechargeable Battery 3.65 V / 2750 mAh / 10.038 Wh
50001508, 1INR19/65, 1INR18650-29E, WWYY
Assembled by Celltech (Zhongshan) Ltd.)
Normal battery service life is expected to be 1–2 years. When the battery is at the end of its service life, the
usage time of the device is reduced. It is recommended to remove the battery if the device is stored for
more than 2 months.