30 Optomed Aurora
6.1 Environmental conditions
The device is intended for use indoors. Do not store or use the device in ambient conditions other than
those prescribed.
USE STORAGE TRANSPORTATION
Temperature +10 °C to +35 °C –20 °C to +35 °C –20 °C to +50 °C
Relative humidity 10 % to 80 % 10 % to 85 % 10 % to 85 %
Atmospheric pressure 800 hPa to 1060 hPa 500 hPa to 1060 hPa 500 hPa to 1060 hPa
NOTE!
THIS INSTRUMENT DOES NOT MEET THE REQUIREMENTS OF ISO 15004-1 FOR STORAGE. DO NOT
STORE THIS INSTRUMENT IN CONDITIONS WHERE THE TEMPERATURE MAY RISE ABOVE +35 °C OR
FALL BELOW -20 °C.
Please also note that electromagnetic compatibility information and recommended separation distances
between portable and mobile RF communications equipment and Optomed Aurora are given in Chapter 6.8.
Protection against harmful ingress of water or particulate matter is defined as IPX0 (Equipment not protected
against the ingress of water).
6.2 Error messages
Optomed Aurora will display error messages to indicate error situations. An error message is always displayed
with an explanatory message providing information about possible actions.
ERROR MESSAGE POSSIBLE ACTION
Optics failure # Please detach the optics module and restart the camera.
Error code # identifies the error and it may be needed by
customer service.
Light source temperature too high This indicates that the light source temperature exceeds
the safety limits. Please wait until the camera cools down.
This may take several minutes.
Image storage not found This indicates that there is a problem with the memory card.
Please contact customer service for more information.
Battery failure The battery is either not properly in place or has been damaged.
Please try reinstalling or replacing the battery and restart.
6. Important information