Shows the flattop correction value. See also SET_CORR_FLAT.
Responses:
CORR_FLAT 000000000000000
SHOW_CRM
Returns the current reading of the count-rate meter.
Response:
$G0000050781096 Current input count rate is 50781 counts/s.
SHOW_CRM_WIDTH
Returns the CRM output pulse width in microsceonds.
Responses:
CRM_WIDT 00000000004.225 Width set to 4.225 us
SHOW_DATA_APPLICATION “string”
If string matches “string1" in a previous SET_DATA_APPLICATION command, then
string2 from that command is returned. The response is a $F record.
Non-USB Execution Errors:
%131138085<CR> The specified string could not be matched.
SHOW_DATE
Returns the day, month, and year of the current date as maintained in the battery-backed-up
real time clock, in the form dddmmmyyy. The day is returned as a 3-digit integer number
from 001 to 031, month as a 3-digit integer number from 001 to 012, and year as a 3-digit
integer number from 000 to 099. See also SET_DATE_START.
Responses:
$N001001088052<CR> Date reported as Jan 1, 1988.
... ...
$N031012099059<CR> Date reported as Dec 31, 1999.
$N001001000036<CR> Date reported as Jan 1, 2000.
... ...
$N031012087056<CR> Date reported as Dec 31, 2087.
SHOW_DATE_START
Returns the day, month and year of the acquisition start date in the form dddmmmyyy. The
day is returned as a 3-digit integer number from 001–031, month as a 3-digit integer number
from 001–012, and year as a 3-digit integer number from 000–099. See also
SET_DATE_START.