Reports the sum of the specified group of spectral data channels. If start_chan and
number_of_chans is not provided, SHOW_INTEGRAL reports the sum of all channels in the
currently selected window that have their ROI flag set.
Responses:
$G0000000000075<CR> Integral reported as 0.
... ...
$G4294967294131<CR> Integral reported as 4294967294.
$G4294967295132<CR> Integral reported as greater than or equal to
4294967295 (maximum reportable value).
SHOW_INTEGRAL_PRESET
Reports the current ROI integral preset value. For more information about the ROI integral
preset, see SET_INTEGRAL_PRESET. See also SHOW_INTEGRAL.
Responses:
$G0000000000075<CR> Integral preset reported as 0.
... ...
$G4294967295132<CR> Integral reported as 4294967295.
SHOW_INTEGRAL_REMAINING
Reports the current ROI integral remaining value. For more information about the ROI
integral remaining, see SET_INTEGRAL_REMAINING. See also SHOW_INTEGRAL.
Responses:
$G0000000000075<CR> Integral remaining reported as 0.
... ...
$G4294967295132<CR> Integral reported as 4294967295.
SHOW_LENGTH_SAMPLE
Reports the number of points in the Insight mode waveform.
Response:
$C01000088 1000 points in the waveform.
SHOW_LFR
Tells whether LFR mode is enabled.
Responses:
$IF LFR mode disabled.
$IT LFR mode enabled.
SHOW_LIVE
Reports the contents of the live-time counter in units of 20 ms (50 ticks/s). See also
CLEAR_COUNTER and SET_LIVE.
108