932502G / 0618 7. FIRMWARE COMMANDS AND RESPONSES
Responses:
$G0000000000075<CR> Live time reported as 0 ticks.
$G0000000001076<CR> Live time reported as 1 tick (20 ms).
... ...
$G4294967295132<CR> Live time reported as 4294967295 ticks (over 23000
days).
SHOW_LIVE_PRESET
Reports the current live-time preset in units of 20 ms (50 ticks/s). See also
CLEAR_PRESETS and SET_LIVE_PRESET.
Responses:
$G0000000000075<CR> Live-time preset reported as disabled.
$G0000000001076<CR> Live-time preset reported as 1 tick.
... ...
$G4294967295132<CR> Live-time preset reported as 4294967295 ticks.
SHOW_LIVE_REMAINING
Reports the current live-time remaining in units of 20 ms (50 ticks/s). See also
SET_LIVE_REMAINING.
Responses:
$G0000000000075<CR> Live-time remaining reported as disabled.
$G0000000001076<CR> Live-time remaining reported as 1 tick.
... ...
$G4294967295132<CR> Live-time remaining reported as 4294967295 ticks.
SHOW_LLD
Shows the lower level discriminator setting. See also SET_LLD.
Responses:
$C00050092 The LLD is 50.
SHOW_MDA [PeakNumber]
Reports current MDA for the MDA Preset selected by PeakNumber. PeakNumber is
assumed to be 0 if not supplied.
Example response:
MDA 0000000000010.7
SHOW_MDA_COEF
Reports the current settings for coeffficients a, b, and c used for the MDA calculation. See
SET_MDA_COEF.
Example Responses:
MDA_COEF 000000000002.71 000000000000000 00000021.700001