DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual 932502G / 0618
Coefficient a = 2.71, b = 0, and c = 21.7.
SHOW_MDA_PRESET [PeakNumber]
Reports MDA preset PeakNumber. If PeakNumber is omitted, 0 is assumed.
Example Response:
MDA_PRES 0000000000025.7
SHOW_MODE
Reports the current mode of operation (PHA, List, or Sample [InSight Virtual
Oscilloscope]). See also SET_MODE_PHA and SET_MODE_SAMPLE.
Responses:
$FPHA<CR> PHA mode.
$FLIS<CR> List mode.
$FSAM<CR> Sample mode (InSight).
SHOW_MODE_ZDT
Reports the current ZDT mode. See SET_MODE_ZDT.
Responses:
MODE_ZDT 00000 Mode 0 selected (normal and corrected).
MODE_ZDT 00001 Mode 1 selected (error and corrected).
SHOW_MONI_LABEL num
Returns with an ASCII string label designation for the SOH parameter for num.
SHOW_MONI_MAX
Returns with the number of SOH values available for viewing.
SHOW_MONI_VALUE num
Returns with an ASCII representation of the value for the SOH parameter.
SHOW_NAI
Indicates whether ENABLE_NAI has been sent. Responses are true and false.
Responses:
$IT<CR> NaI mode enabled.
$IF<CR> NaI mode disabled.
SHOW_NEXT
Used in conjunction with the SHOW_ROI command, SHOW_NEXT reports the next contin-
uous group of channels that have the ROI flag set. The response is of the form:
$Dsssssnnnnnccc<CR>