DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual 932502G / 0618
largest count is reported if more that one channel contains the largest number of counts.
Channel 16383 is the highest numbered channel in any device.
Responses:
$C00000087<CR> Maximum count was found in channel 0 or no ROI channels
were found.
$C00001088<CR> Maximum count was found in channel 1.
... ...
$C16382107<CR> Maximum count was found in channel 16382.
$C16383108<CR> Maximum count was found in channel 16383.
SHOW_PEAK_PRESET
Reports the value of the ROI peak preset. See SET_PEAK_PRESET for information about
the ROI peak preset.
Responses:
$G0000000000075<CR> Peak preset disabled.
$G0000000001076<CR> Peak preset set to 1 count.
$G2147483646120<CR> Peak preset set to 2147483646 counts.
$G2147483647121<CR> Peak preset set to 2147483647 counts.
SHOW_PUR
Returns the status of the Pile-Up Rejection Circuitry.
Responses:
$IT<CR> PURenabled.
$IF<CR> PUR disabled.
SHOW_PZ
Displays the PZ setting. See also SET_PZ.
Responses:
PZ_ENA 002200 Pole zero enabled and set to 2200.
PZ_DIS 001000 Pole zero disabled and set to 1000.
SHOW_PZ_AUTOMATIC
Shows the state of the automatic PZ. See also SET_PZ_AUTOMATIC.
Responses:
$IT<CR> Pole zero enabled.
$IF<CR> Pole zero disabled (TRP).