DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual 932502G / 0618
Returns the SCA output settings in the form SCA_OUTP www.ww a.
Responses:
SCA_OUTP 00000000004.525 000000000000001 SCA pulse width is 4.525 us and
always active.
SHOW_SEGMENT
Returns the current segment number. Always 1 for the DSPEC 50.
SHOW_SHAP_FLAT
Reports the width of the flattop in ìs. See also SET_SHAP_FLAT.
Responses:
SHAP_FLAT 0000000000001.2
SHOW_SHAP_RISE
Displays the rise-time setting in ìs. See also SET_SHAP_RISE.
Responses:
SHAP_RISE 0000000000003.2
SHOW_SHUTDOWN
Shows the type of HV shutdown selected.
Responses:
$FORT ORTEC mode selected.
$FTTL TTL mode selected.
$FSM1 SMART-1 mode selected
SHOW_SHUT_ACTUAL
Returns the current status of the bias remote shutdown input signal. This command is valid
whether the bias supply is turned on or off.
Responses:
$IT Shutdown is active (supply is shut down or can’t be
turned on).
$IF Shutdown is inactive (supply is on or can be turned on).
SHOW_SNUM
Responds with a $F record indicating the serial number of the DSPEC 50
Response:
$F100 Serial number = 100.
SHOW_STATUS
Returns system status information in the following format:
$Mllllllllllttttttttttaaaaahhhhhccc<CR>
114