DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual 932502G / 0618
SHOW_UNCERTAINTY
Returns the current value of the uncertainty for the peak in the uncertainty preset. See also
SET_UNCERTAINTY.
Responses:
UNCE 0000000000008.5 Uncertainty of the peak is 8.5%.
SHOW_UNCERTAINTY_PRESET
Returns the current uncertainty preset setting. See also SET_UNCERTAINTY_PRESET.
Responses:
UNCE_PRES 000000000000000 No preset.
UNCE_PRES 0000000000008.5 Preset set to 8.5%.
SHOW_VERSION
Reports the DSPEC 50 firmware version number in the form
Fmmmm-vvv<CR>
where mmmm is a 4-character model designator and vvv is a 3-character version designator.
Example Responses:
$FDS50–003<CR> Model DSPEC 50 firmware version 3 reported.
SHOW_WINDOW
Reports the start channel and number of channels in the window of interest, in the form
$Dxxxxxyyyyyccc<CR>
where xxxxx is the start channel (0–16383) and yyyyy is the number of channels (1–16384).
See SET_WINDOW for more information about the window of interest.
Example Responses:
$D0000016384094<CR> Window of interest reported as starting at channel 0
and continuing for 16384 channels.
$D0000008192092<CR> Window of interest reported as starting at channel 0
and continuing for 8192 channels.
$D0819208192112<CR> Window of interest reported as starting at channel
8192 and continuing for 8192 channels.
SHOW_ZDT
Reports if ZDT mode is enabled.
Responses:
$IF ZDT mode disabled.
$IT ZDT mode enabled.
SHOW_ZERO_ADJ
Returns the value of the zero offset for the zero stabilizer.
118