Digital Gamma-Ray Spectrometer User’s Manual 932502G / 0618
DISABLE_NAI
Switches the stabilizers to Ge mode. See ENABLE_NAI.
DISABLE_OPTI_NEED
Disables the reporting of an error when the DSPEC 50 has not been optimized.
DISABLE_PUR
Disables the Pile-Up Rejection Circuitry.
DISABLE_PZ_AUTOMATIC
Disables the automatic pole zero mode. See also ENABLE_PZ_AUTOMATIC.
DISABLE_PZ_NEED
Disables the reporting of an error when the DSPEC 50 has not been pole zeroed.
DISABLE_REMOTE
No function in the DSPEC 50; included for backward compatibility.
DISABLE_RENH
Returns the DSPEC 50 to normal operation if the Resolution Enhancer is enabled. See
ENABLE_RENH.
DISABLE_THRESHOLD_FMAN
Disables manual adjustment of the fast channel threshold.
DISABLE_THRESHOLD_AUTOMATIC
DISABLE_THRESHOLD_SAMPLE
DISABLE_TRIGGER_SPECIAL
No function in the DSPEC 50; included for backward compatibility.
DISABLE_TRIGGER
Disables the data acquisition trigger that was enabled by the ENABLE_TRIGGER
command. Cancels an ENABLE_TRIGGER_REP if it is issued before the acquisition starts.
See ENABLE_TRIGGER, TRIGGER, and SHOW_TRIGGER.
DISABLE_ZERO_STABILIZATION
Stops stabilization of the zero peak while data is being acquired. The zero stabilization
adjustment is held at its current value until either zero stabilization is reenabled with the
ENABLE_ZERO_STABILIZATION command or reinitialized with the
INITIALIZE_ZERO_ STABILIZATION, SET_ZERO_CHANNEL or SET_ZERO_WIDTH
commands. See also SHOW_ZERO_STABILIZATION.