DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual 932502G / 0618
LIST_FAST_COIN x
Lists the valid settings for fast coincidence delay and window. If x=0 then the valid
Window settings are returned. If x=1 then the valid Delay settings are returned.
Response:
FAST_COIN 0 6.375 1 255 Valid Window settings are 0 to 6.375 uS with up to 255
steps of 0.025 uS starting with 0.025 uS minimum.
FAST_COIN -6.375 6.375 0 510 Valid Delay settings are -6.375 to 6.375 uS with up to
510 steps of 0.025 uS.
LIST_GAIN_COARSE
Lists the valid coarse-gain settings.
Response:
GAIN_COAR 1 2 4 8 16 32 64 128
LIST_GAIN_CONV
Lists legal conversion gain settings.
Response:
GAIN_CONV 512 1024 2048 4096 8192 16384
LIST_GAIN_FINE
Lists the valid fine-gain settings.
Response:
GAIN_FINE 0.45 1.0 88437 196607 0.45 to 1 in 108171 steps.
LIST_HV
Lists the valid high-voltage settings.
Response:
HV 0 5000 0 5000 0 to 5000 in 5000 steps.
LIST_INHIBIT_WIDTH
Lists the valid inhibit width settings.
Response:
INHI_WIDT 0 200 0 8000 0 to 200 uS with up to 8000 steps of 0.025 uS
LIST_MODE_ZDT
Reports the legal ZDT-mode settings. Each item in the list can be selected by sending the
SET_MODE_ZDT command with a mode number. The mode number is implied by the
location of the mode in the list. For example, the first item in the list had a mode number
of 0. To select the first item, send SET_MODE_ZDT 0. The second item has a mode number
of 1, and so on.
Response:
88