22 Notes on EMC (Electromagnetic Compatibility)
ICS Impulse USB is part of a medical electrical system and is thus subject to special safety precautions. For this reason,
the installation and operating instructions provided in this document should be followed closely.
Portable and mobile high-frequency communication devices, such as mobile phones, may interfere with the func-
tioning of ICS Impulse USB.
Guidanceand manufacturer's declaration - electromagnetic emissionsfor all equipment and systems
ICSImpulse USBis intended for use in the electromagnetic environment specified below. The user of ICS Impulse USBshould ensure that itis used in
such an environment.
Emissions test Compliance Electromagneticenvironment - guidance
RF emissions
CISPR 11
Group 1 ICS Impulse USB uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any inter-
ference in nearby electronic equipment.
RF emissions
CISPR 11
Class A ICS Impulse USB is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Immunity - All Equipment and Systems
The ICS Impulse USB System was not tested for immunity to electromagnetic disturbances.
128 Otometrics - ICS Impulse USB
22 Notes on EMC (Electromagnetic Compatibility)