How to add anti-freeze agent:
Preparing the anti-freeze solution: Mix 1 litre of anti-freeze with 0.4 litre of
w
ater.
Carefully add anti-freeze solution to the collecting bucket (the pipe housing
does not have to be removed for this).
Please note: Add anti-freeze agent only as an aqueous solution (add 40 % of
water)! Never use neat! (POWERCOOL is hygroscopic.)
Note on disposing of anti-freeze solution
Typically, the anti-freeze solution of a single precipitation gauge may be disposed
of into the public sewage system. However, observe all applicable local regulations.
Please refer to your local authority responsible for questions regarding disposal
and/or contact Thermochema GmbH.
Note on winter operation without anti-freeze
Basically, the unit may be operated in the winter season without adding anti-
freeze solution. However, this mandatorily requires the collecting bucket to be
checked and emptied in regular intervals!
Please note: When the accumulated precipitation freezes that has
filled the collecting bucket up to 80 %, this may result in perma-
nent damage to the weighing mechanism.
Note on third-party anti-freeze solution products
Basically, third-party products may be used provided the following requirements
are met:
Good solubility in water, low-density (i.e. the anti-freeze agent does not settle
below the water).
Low evaporation (do not use methanol).
Little corrosive effect on aluminium and stainless steel.
Low freezing point even with high collecting bucket level.
Low hygroscopic properties (absorption of moisture from ambient air, which
would affect the measurement results).
Note chemical compatibility with collecting bucket (ASA).
No gumming after use in open containers for several months.
7.3 Performing the visual inspection
Check smooth movement of the collecting bucket in all directions at the lower
edge of the bucket orifice. The collecting bucket upper edge may not contact
the pipe housing!
If present, carefully remove any contamination (e.g. insects, insect nests, spider
webs, etc.), and icing.
7.4 Additional checks in cases of defect
Is the collecting bucket distorted?
Is the collecting bucket correctly positioned on the bucket support?
Is the pipe housing correctly positioned and not damaged?
Is the weighing mechanism contacting other components, e.g. caused by
contamination?
Is the red LED flashing (refer to Fig. 18)?
Use Chapter 7.5 to isolate the failure.
If in doubt, perform a Guided Accuracy Test as described in Chapter 7.6.