EasyManuals Logo

PerkinElmer Lambda 465 Installation Instructions

PerkinElmer Lambda 465
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
09931238D
23
8. The Setup Single Cell Peltier window will be shown. Set parameters according to the
experiment conditions.
Temperature Control
a. Temperature (°C): Enter the preset temperature for the experiment.
NOTE:
Starting the temperature of experiment needs to be set up in the method window.
b. Return to temperature (°C): Enter the returning temperature after the experiment is finished.
c. Error Range (°C)
Enter the temperature tolerance between the sampling temperature and monitored
temperature. The measurement will start when the temperature tolerance reaches within the
set Error Range.
Example: Set the starting temperature as 20°C and error range as 1°C then, the monitoring
temperature is set as the temperature of cell block. In this case, the measurement will start
when the temperature of the cell block reaches between 19°C and 21°C.
NOTE:
When selecting probe for Monitor in Collect Temperature box, the slower the real heating speed may be,
the smaller the set Error range (°C) is.
Collect Temperature
a. Monitor: Select which temperature will be monitored in the display panel.
- Block: The temperature of Cell Block
- Probe 1: The temperature of Probe 1
- Probe 2: The temaperature of Probe 2
Example: If Block is selected, measurement will start when the temperature of the cell block
reaches the setting temperature.
Temperature Display
Select which temperature will be displayed on the panel: Block, Probe 1 or Probe 2.
The selected temperature is only displayed in real time.

Other manuals for PerkinElmer Lambda 465

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda 465 and is the answer not in the manual?

PerkinElmer Lambda 465 Specifications

General IconGeneral
BrandPerkinElmer
ModelLambda 465
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals