Page #12 background imageLoading...
Page #12 background image
09931238D
12
Peltier controller auto tuning setup
Peltier controller can be compatible with various peltier cell holders (Single or Multi) so whenever you
exchange the existing peltier cell holder to the other one, you should proceed “AUTO TUNING” set up to
minimize the temperature fluctuation at the target temperature
1. Push the Manual button on the control panel of the peltier controller then button light is on.
2. Set any SP temperature between 30~37°C using the button.
3. Press and hold the SET/ENT button until shown “G.AT”.