EasyManua.ls Logo

Permobil Explorer Mini - Electromagnetic Immunity

Permobil Explorer Mini
142 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Explorer Mini Electromagnetic compatibility
8.2 Electromagnetic immunity
Guidance and manufacturer´s declaration
The Explorer Mini and the battery charger (PF2401A5SL) are intended for use in the electromagnetic environment specified
below. The customer or the user of the Explorer Mini and the battery charger (PF2401A5SL) should assure that they are used in
such an environment
Explorer Mini
Immunity test
Test level IEC 60601 Compliance level Electromagnetic environment guidance
Electrostatic discharge
(ESD)IEC 61000-4-2
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m 30 A/m
If image distortion occurs, it may be necessary to
position the Explorer Mini further from sources of
power frequency magnetic fields or to install
magnetic shielding. The power frequency
magnetic field should be measured in the intended
installation location to assure that it is sufficiently
low.
Battery charger (PF2401A5SL)
Immunity test
Test level IEC 60601 Compliance level Electromagnetic environment guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±2 kV, ±4 kV, ±6 kV,
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.