17
SimplyGo User Manual
Item DescrIptIon
4 Amount of full charge currently in the battery�
5 External power is connected�
6 Software version in the device�
3� Press the Power button a second time to place the device in
operating mode� This second press helps prevent unwanted starts
or inadvertent changes in oxygen ow� The Introductory screen
momentarily appears, shown at left below, followed by the nal
operating mode screen, shown at right�
Every time it starts up, the device starts producing oxygen to meet
the specied setting� It may take up to ten minutes to reach the
specied purity levels for the setting� It is safe to begin breathing
from the SimplyGo device immediately� Breathing from the device
will cause it to reach oxygen purity levels faster than ten minutes�
The dot in the lower right corner of the LCD display is the pulse
symbol� It appears whenever you take a breath and the device
delivers a pulse of oxygen�
Note: When it is turned on, the device will automatically start at the ow rate
settingusedwhenthedevicewaslastturnedo.Asaprecaution,each
time you start the device, verify the ow setting�
4� To set the device’s ow setting, press either the + or – button�
To increase the ow setting, press the + button� To decrease the ow
setting, press the – button�
The device can be set to 1, 1�5, 2, 2�5, 3, 3�5, 4, 4�5, 5, 5�5, and 6 in
Pulse or Sleep Mode� It can be set to �5, 1, 1�5, and 2 in Continuous
FlowMode.
Warning: It is very important to set your device to your prescribed level of
oxygen ow� Do not increase or decrease your ow rate from your
prescribed level until you rst consult with your physician�