10 /20
Use, maintenance and calibration
M0147E
6.1.2 RESETTING THE RESET TOTAL
The reset total resetting operation can only be performed after resetting the
partial register. The reset total can in fact be reset by pressing the reset key at
length while the display screen shows reset total as on the following display
page:
Schematically, the steps to be taken are:
1
Wait for the display to show normal standby display page (with total only dis-
played)
2 Press the reset key quickly
3 The meter starts to reset the partial
4 While the display page showing the reset total is displayed
Press the reset key again for at least 1 second
5 The display screen again shows all the segments of the display followed by all
the switched-off segments and finally shows the display page where the reset
Reset Total is shown.
6.2 DISPENSING IN FLOW RATE MODE
FOREWORD
It is possible to dispense, displaying at the same time:
- the dispensed partial
- the Flow Rate in [Partial Unit / minute] as shown on the following
display page:
Procedure for
entering this
mode:
- wait for the meter to go to Standby, meaning the display screen shows Total only
- quickly press the CAL key.
- Start dispensing
The flow rate is updated every 0.7 seconds. Consequently, the display could be relatively
unstable at lower flow rates. The higher the flow rate, the more stable the displayed value.
ATTENTION
The flow rate is measured with reference to the unit of measurement of the Par-
tial. For this reason, in case of the unit of measurement of the Partial and Total
being different, as in the example shown below, it should be remembered that
the indicated flow rate relates to the unit of measurement of the partial. In the
example shown, the flow rate is expressed in Qts/min.
The word “Gal” remaining alongside the flow rate refers to the register of
the Totals ( Reset or NON Reset) which are again displayed when exiting
from the flow rate reading mode.
To return to “Normal” mode, press the CAL key again. If one of the
two keys RESET or CAL is accidentally pressed during the count,
this will have no effect.
.
ATTENTION Even though in this mode they are not displayed, both the Reset Total
and the General Total (Total) increase. Their value can be checked
after dispensing has terminated, returning to “Normal” mode, by
quickly pressing CAL.