Use, maintenance and calibration
M0147E
Gal
Qts
Pts
Gal
RESET
RESET
RESET
4
CALCAL
CAL
By pressing the CAL key at length, the new settings will be stored, the METER will
pass through the start cycle and will then be ready to dispense in the set units.
ATTENTION The Reset Total and Total registers will be automatically changed to the
new unit of measurement.
NO new calibration is required after changing the Unit of Measurement.
9 MAINTENANCE
9.1 CHANGE BATTERY
FOREWORD
The METER has been designed to require a minimum amount of maintenance.
The only maintenance jobs required are:
• Battery change – necessary when the batteries have run down
• Cleaning the measurement chamber. This may be necessary due to the par-
ticular nature of the dispensed fluids or due to the presence of solid particles
following bad filtering
BATTERY
REPLACEMENT
The METER is complete with 2 x 1.5 V. alkaline batteries SIZE N.
WARNING K600 should be installed in a position allowing the batteries to be re-
placed without removing it from the system.
K600 features two low-battery alarm levels:
1 When the battery charge falls below the first level on the LCD, the fixed battery
symbol appears. In this condition, K600 continues to operate correctly, but
the fixed icon warns the user that it is ADVISABLE to change the batteries.
2 If K600 operation continues without changing the batteries, the second battery
alarm level will be reached which will prevent operation. In this condition the bat-
tery icon starts to flash and is the only one to remain visible on the LCD.
To change the
batteries, with
reference to
the exploded
diagram posi-
tions, proceed as
follows
1
Press RESET to update all the totals
2
Unscrew the battery cap (pos.8)
3
Remove the old batteries
4
Place the new batteries in the same position as the old ones, making sure the
positive pole is positioned as indicated on the cover (pos. 9). Re-tighten the bat-
tery cap, making sure the seal (pos.1) are correctly positioned.
5
The METER will switch on automatically and normal operation can be resumed.
The METER will display the same Reset Total, the same Total and the same Partial indicated before
the batteries were changed. After changing the batteries, the meter does not need calibrating again.
ATTENTION Do not discard the old batteries in the environment. Refer to local
disposal regulations.